ACCU-CHEK ® SPIRIT
Report
- Report Number
- 2183996-2013-00895
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 14, 2013
- Report Date
- January 23, 2014
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
AS THE PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION AND THE PRODUCTION DATA COMPLY WITH THE SPECIFICATION, THE COMPLAINT COULD NOT BE REPLICATED. PRODUCTION REPORTS WERE REVIEWED.
THE COMPLAINT CANNOT BE VERIFIED DUE TO CARELESS HANDLING OF THE PRODUCT. THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY. THEREFORE, MOISTURE MAY ENTER THE INSULIN PUMP AND DESTROY THE PUMP ELECTRONICS. THE FUNCTIONALITY OF THE BUTTONS WAS TESTED SUCCESSFULLY.
PATIENT REPORTED THE UP AND DOWN BUTTONS ON THE INFUSION DEVICE ARE NON- FUNCTIONAL. THE INFUSION DEVICE WAS REPLACED AND REQUESTED FOR EVALUATION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND THE PATIENT DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223768 | ACCU-CHEK ® SPIRIT | INSULIN INFUSION PUMP | LZG | ROCHE DIABETES CARE AG | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 012 YR |