FDA Adverse Event Death Summary report: N

ON-Q C-BLOC: 600 ML, 2-14 ML/HR SELECT-A-FLOW

MDR report key: 3121602 · Received May 16, 2013

Report

Report Number
2026095-2013-00092
Event Type
Death
Date Received
May 16, 2013
Report Date
April 13, 2013
Manufacturer
I-FLOW LLC
Product Code
MEB
PMA / PMN Number
K063530
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE SAMPLE WILL NOT BE RETURNED FOR AN EVAL AND INVESTIGATION. RESULTS: THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED. CONCLUSIONS: AT THIS TIME, NO CONCLUSION CAN BE DRAWN. THIS COMPLAINT IS STILL UNDER INVESTIGATION . WHEN THE INVESTIGATION IS COMPLETED, I-FLOW WILL SUBMIT A FOLLOW-UP REPORT. I-FLOW HAS ATTEMPTED TO CONTACT THE DOCTOR SEVERAL TIMES TO FURTHER INVESTIGATE THIS INCIDENT WITH NO RESPONSE. A MESSAGE WAS LEFT WITH THE PHYSICIAN'S SURGICAL ASSISTANT TO HAVE THE DOCTOR RETURN OUR CALLS. INFORMATION FROM THIS INCIDENT WILL BE INCLUDED IN OUR PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEM. ADDITIONAL INVESTIGATION MAY ARISE FROM ONGOING ANALYSIS, TREND INFORMATION, OR OTHER ANALYSIS AS APPROPRIATE.

Description of Event or Problem · 1

DRUG/DILUENT: 0.5% MARCAINE, FILL VOLUME: 400 ML, FLOW RATE: 14.0 ML/HR, PROCEDURE: DONOR SITE BURN GRAFTING, CATHPLACE: UNK. AN UNIDENTIFIED HOSPITAL PERSONNEL CONTACTED AN I-FLOW SALES REP TO REPORT AN INCIDENT OF (B)(6) BURN PT. THE PT RECEIVED A CB6004 (600 ML) PUMP INSTEAD OF A P400X4D (400 ML) PUMP AFTER SKIN GRAFT SURGERY. THE CB6004 (600 ML) PUMP USED ON THE PT AND WAS FILLED TO 400 ML WITH SAF SET AT 14 ML/HR. IT WAS NOTICED BY A MEDICAL RESIDENT THAT THE VOLUME OF THE PUMP WAS DEPLETED IN 28 HOURS. THE PT WAS GIVEN LIPIDS WHEN THE INCIDENT WAS DISCOVERED. THIS PT CODED AND COULD NOT BE RESUSCITATED. DATE OF DEATH IS UNK AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
217747 ON-Q C-BLOC: 600 ML, 2-14 ML/HR SELECT-A-FLOW ELASTOMERIC PUMP MEB I-FLOW LLC 101347600 UNK

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death