FDA Adverse Event Injury Summary report: N

VA LOCKSCR Ø2.4 SELF-TAP L18 TAN

MDR report key: 3121542 · Received May 21, 2013

Report

Report Number
2520274-2013-02732
Event Type
Injury
Date Received
May 21, 2013
Report Date
April 23, 2013
Manufacturer
SYNTHES USA
Product Code
HWC
PMA / PMN Number
K103243
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.

Additional Manufacturer Narrative · 1

THIS DEVICE IS FOR TREATMENT, NOT DIAGNOSIS. THE MEASURABLE DIMENSIONS OF THE BROKEN VA LOCKING SCREWS STARDRIVE® WERE CHECKED AND FOUND TO BE IN COMPLIANCE WITH THE TECHNICAL DRAWINGS AND AO/ASIF SPECIFICATION. THE LOT NUMBERS ARE NOT KNOWN, THEREFORE THE RAW-MATERIAL TESTING CERTIFICATE AND THE MANUFACTURING DOCUMENTS CANNOT BE REVIEWED AND THE PRESENT COMPLAINT CANNOT BE FULLY ANALYSED. WE CAN ONLY CONFIRM THAT OUR VA LOCKING SCREWS MADE FROM TITANIUM ALLOY ARE MANUFACTURED IN COMPLIANCE WITH THE AO/ASIF SPECIFICATION AND WITH THE INTERNATIONAL STANDARD ISO 5832-11. UNFORTUNATELY THE INFORMATION GIVEN DID NOT PROVIDE SUFFICIENT EVIDENCE FOR AN EXACT DETERMINATION OF THE FAILURE REASON. NEVERTHELESS, A MANUFACTURING RELATED CONDITION CAN BE EXCLUDED.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON EXPERIENCED BREAKAGE OF THREE 2.4 SCREWS IN A DISTAL RADIUS PLATE. 5 SCREWS WERE IN THE DISTAL ROW. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224753 VA LOCKSCR Ø2.4 SELF-TAP L18 TAN HWC SYNTHES USA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention