VA LOCKSCR Ø2.4 SELF-TAP L16 TAN
Report
- Report Number
- 2520274-2013-02731
- Event Type
- Injury
- Date Received
- May 21, 2013
- Report Date
- April 23, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- HWC
- PMA / PMN Number
- K103243
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.
ADDITIONAL NARRATIVE: AN INVESTIGATION WAS CONDUCTED AND IT STATES THAT THE MEASURABLE DIMENSIONS OF THE BROKEN VA LOCKING SCREWS STARDRIVE WERE CHECKED AND FOUND TO BE IN COMPLIANCE WITH THE TECHNICAL DRAWINGS AND AO/ASIF SPECIFICATION. THE LOT NUMBERS ARE NOT KNOWN, THEREFORE THE RAW-MATERIAL TESTING CERTIFICATE AND THE MANUFACTURING DOCUMENTS CANNOT BE REVIEWED AND THE PRESENT COMPLAINT CANNOT BE FULLY ANALYZED. WE CAN ONLY CONFIRM THAT OUR VA LOCKING SCREWS MADE FROM TITANIUM ALLOY ARE MANUFACTURED IN COMPLIANCE WITH THE AO/ASIF SPECIFICATION. UNFORTUNATELY, THE INFORMATION GIVEN DID NOT PROVIDE SUFFICIENT EVIDENCE FOR AN EXACT DETERMINATION OF THE FAILURE REASON. NEVERTHELESS, A MANUFACTURING RELATED CONDITION CAN BE EXCLUDED.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE ADDITIONAL EVALUATION REPORTS AS RECEIVED CONDITION TO BE A BREAKAGE OF THREE VA LOCKING SCREWS. THE MEASURABLE DIMENSIONS OF THE BROKEN VA LOCKING SCREWS STARDRIVE WERE CHECKED AND FOUND TO BE IN COMPLIANCE WITH THE TECHNICAL DRAWINGS AND AO/ASIF SPECIFICATION. THE LOT NUMBERS ARE NOT KNOWN, THEREFORE THE RAW-MATERIAL TESTING CERTIFICATE AND THE MANUFACTURING DOCUMENTS CANNOT BE REVIEWED AND THE PRESENT COMPLAINT CANNOT BE FULLY ANALYZED. WE CAN ONLY CONFIRM THAT OUR VA LOCKING SCREWS MADE FROM TITANIUM ALLOY ARE MANUFACTURED IN COMPLIANCE WITH THE AO/ASIF SPECIFICATION AND WITH THE INTERNATIONAL STANDARD ISO 5832-11. UNFORTUNATELY THE INFORMATION GIVEN DID NOT PROVIDE SUFFICIENT EVIDENCE FOR AN EXACT DETERMINATION OF THE FAILURE REASON. NEVERTHELESS, A MANUFACTURING RELATED CONDITION CAN BE EXCLUDED. THE DISPOSITION OF THIS EVALUATION IS DEEMED INVALID.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON EXPERIENCED BREAKAGE OF THREE 2.4 SCREWS IN A DISTAL RADIUS PLATE. 5 SCREWS WERE IN THE DISTAL ROW. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
THIS IS REPORT 1 OF 3 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224711 | VA LOCKSCR Ø2.4 SELF-TAP L16 TAN | HWC | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |