OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-12214
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Report Date
- April 24, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (07/29/2013). THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 5/20/2013 AND 7/24/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF LIFESCAN RECEIVES THE PRODUCT LIFESCAN WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING DOES NOT TURN OFF. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222503 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3418762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |