FDA Adverse Event Malfunction Summary report: N

COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER

MDR report key: 3121264 · Received May 20, 2013

Report

Report Number
1061932-2013-00917
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 22, 2013
Report Date
April 22, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED FLUID LEAKED AT THE BLOOD SAMPLING VALVE (BSV). THE FSE REPLACED THE BSV ACTUATOR, REROUTED AND REFITTED BSV TUBING, AND RESOLVED THE FLUID LEAK ISSUE. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY TWENTY (20) MILLILITERS OF CLENZ CLEANER SOLUTION LEAKED FROM THE MANUAL PROBE AND OUTSIDE THE INSTRUMENT INVOLVING THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. THE FACILITY HAS AN EXPOSURE CONTROL PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.