FDA Adverse Event
Malfunction
Summary report: N
RESERVOIR 1.8ML
MDR report key: 3121176
·
Received May 20, 2013
Report
- Report Number
- 3004209178-2013-93770
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 22, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
INSPECTED THREE RANDOM SEALED RESERVOIRS AND ONE OPEN/USED RESERVOIR. MANUAL PRIME AND HIGH-PRESSURE TEST WERE PERFORMED AND ALL THE RESERVOIRS PASSED THE TESTS. NO LEAKS OR DEFECTS IN THE O-RINGS WERE OBSERVED DURING ANALYSIS AND ALL THE RESERVOIRS PASSED PER SPECIFICATIONS.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED PAST THE O-RINGS AND INTO THE RESERVOIR COMPARTMENT. THE BLOOD GLUCOSE READING WAS 311MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222828 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A | H8486688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |