FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3121156 · Received May 20, 2013

Report

Report Number
3004209178-2013-93775
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 30, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO OCCLUDED ANOMALY WAS OBSERVED DURING ANALYSIS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED PAST THE O-RINGS. THE BLOOD GLUCOSE READING WAS 254MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221377 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 46 YR