FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3121150
·
Received May 20, 2013
Report
- Report Number
- 3004209178-2013-93762
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 27, 2013
- Report Date
- May 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH AN ALARM DURING THE BASIC OCCLUSION TEST AS A RESULT OF A MISSING MOTOR SUPPORT DISK. THE DEVICE HAD A CRACKED DISPLAY WINDOW CORNER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 540MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT DEVICE ALARMED. THE CALLER STATED THAT THE DRIVE SUPPORT CAP APPEARS TO BE SLIGHTLY INDENTED. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 221752 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |