VENTED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2013-00407
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Report Date
- April 23, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED A BREAK IN THE WATER FEEDSET TUBE WHERE IT CONNECTS TO THE CHAMBER DOME. THE SURFACE OF THE BREAK WAS ROUGH. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 120503. CONCLUSION: THE DAMAGE APPEARS TO BE THE RESULT OF THE TUBE BEING PULLED AWAY FROM THE DOME, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. THE CUSTOMER HAD REPORTED THAT THE FEEDSET TUBE HAD BECOME SEPARATED FROM THE WATER BAG SPIKE BUT THIS WAS NOT THE CASE WITH THE RETURNED COMPLAINT DEVICE. IT IS POSSIBLE THAT THE CUSTOMER HAD EXPERIENCED ISSUES WITH THE BAG SPIKE DETACHING FROM THE FEEDSET TUBE AS A RESULT OF THE USER PULLING ON THE TUBING TO REMOVE THE BAG SPIKE, RATHER THAN GRASPING THE SPIKE ITSELF. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT THEY WERE HAVING PROBLEMS WITH THE CONNECTION OF THE BAG SPIKE AND THE WATER FEEDSET ON THE MR290V HUMIDIFICATION CHAMBERS. THEY STATED THAT THE BAG SPIKE WAS REMAINING IN THE WATER BAG FOLLOWING REMOVAL OF THE WATER FEEDSET. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222240 | VENTED HUMIDIFICATION CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 120503 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |