FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3120962 · Received May 20, 2013

Report

Report Number
3004209178-2013-93796
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 25, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO ERROR ALARM NOTED ON ALARM HISTORY SCREEN. MISSING END CAP STICKER. MOTOR WAS RECEIVED WITH CORRODED MOTOR HOME SWITCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE DRIVE SUPPORT CAP IS PROTRUDING. CUSTOMER'S BLOOD GLUCOSE READING IS 153 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP HAS ALARMED MOTOR ERROR NUMEROUS TIMES. CUSTOMER COULD NOT EXIT OUT OF THE REWIND LOOP. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222167 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 23 YR