FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3120962
·
Received May 20, 2013
Report
- Report Number
- 3004209178-2013-93796
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO ERROR ALARM NOTED ON ALARM HISTORY SCREEN. MISSING END CAP STICKER. MOTOR WAS RECEIVED WITH CORRODED MOTOR HOME SWITCH.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE DRIVE SUPPORT CAP IS PROTRUDING. CUSTOMER'S BLOOD GLUCOSE READING IS 153 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP HAS ALARMED MOTOR ERROR NUMEROUS TIMES. CUSTOMER COULD NOT EXIT OUT OF THE REWIND LOOP. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222167 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |