FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3120918
·
Received May 20, 2013
Report
- Report Number
- 3004209178-2013-93704
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 25, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO KEYPAD TRACE. NUMBERS DO NOT RAMP UP ON THEIR OWN OR SCROLL BAR MOVING WITHOUT INPUT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP HAS A KEYPAD ISSUE. CUSTOMER STATED THAT THE BLOOD GLUCOSE READINGS HAVE BEEN HIGH. CUSTOMER'S BLOOD GLUCOSE READING IS 415 MG/DL. CUSTOMER TREATED WITH INSULIN PUMP, WHICH IS NOT GOING WELL. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222226 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |