FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3120918 · Received May 20, 2013

Report

Report Number
3004209178-2013-93704
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO KEYPAD TRACE. NUMBERS DO NOT RAMP UP ON THEIR OWN OR SCROLL BAR MOVING WITHOUT INPUT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAS A KEYPAD ISSUE. CUSTOMER STATED THAT THE BLOOD GLUCOSE READINGS HAVE BEEN HIGH. CUSTOMER'S BLOOD GLUCOSE READING IS 415 MG/DL. CUSTOMER TREATED WITH INSULIN PUMP, WHICH IS NOT GOING WELL. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222226 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR