FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3120897 · Received May 20, 2013

Report

Report Number
1061932-2013-00877
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER FOUND A SMALL TEAR IN THE ASPIRATION LINE OF THE NEEDLE CARTRIDGE. THE CUSTOMER REPLACED THE NEEDLE CARTRIDGE RESOLVING THE LEAK. THE ANALYZER IS OPERATIONAL. BECKMAN COULTER SERVICE DID NOT GO ONSITE FOR THIS EVENT. FAILURE MODE OF THE LEAK WAS RELATED TO THE NEEDLE CARTRIDGE TUBING. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT APPROXIMATELY 20 ML OF A DILUTED BLOODY LIQUID LEAKED FROM THE NEEDLE CARTRIDGE AREA IN THE COULTER LH 780 HEMATOLOGY ANALYZER. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE MATERIAL SAFETY DATA SHEET (MSDS WAS NOT REVIEWED. THERE IS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE AT THE FACILITY. THE OPERATOR WAS WEARING A LABORATORY COAT AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. THERE WAS NO IMPACT TO PATIENT RESULTS. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.