COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Report
- Report Number
- 1061932-2013-00875
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 25, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE VERIFIED THAT THE INSTRUMENTS BACKWASH CUP WAS NOT RETRACTING COMPLETELY. UPON FURTHER INSPECTION, THE FSE DISCOVERED THAT THE PROBE WAS BENT AND THERE WAS ALSO A LOOSE BLOOD SAMPLING VALVE (BSV) KNOB. AFTER BENDING THE PROBE BACK INTO PLACE AND TIGHTENING THE BSV KNOB THE BACK WASH CUP RETRACTED COMPLETELY AND THE LEAK STOPPED. THE FSE DOCUMENTED THAT THE LEAK WAS COMING FROM THE BSV DUE TO THE LOOSE BSV KNOB. THE FSE VERIFIED THE INSTRUMENT. FAILURE MODE FOR THE LEAK WAS LOOSE BSV KNOB. THE BACKWASH CUP WAS NOT RETRACTING COMPLETELY DUE TO THE PROBE BEING BENT. (B)(4).
THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THE BACKWASH CUP WAS NOT RETRACTING ALL THE WAY UP AND THERE WAS APPROXIMATELY 1 ML LEAK OF CLEAR LIQUID CONTAINED WITHIN THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE INSTRUMENT WAS RUNNING A STARTUP WHEN THE LEAK WAS OBSERVED. THERE WAS NO ERROR MESSAGE GENERATED IN CONNECTION WITH THIS INCIDENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THERE IS AN EXPOSURE CONTROL OR RISK MANAGEMENT IN PLACE IN THE FACILITY. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES, GOGGLES, AND A LABORATORY COAT. THERE WAS NO EXPOSURE TO ANY OPEN WOUNDS OR MUCOUS MEMBRANES. NO PATIENT SAMPLES WERE AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222075 | COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |