FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ?

MDR report key: 3120838 · Received May 20, 2013

Report

Report Number
2134265-2013-03422
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 22, 2013
Report Date
April 22, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE DEVICE WAS RECEIVED IN TWO SECTIONS AS A RESULT OF A BREAK IN THE HYPOTUBE. THE BREAK WAS LOCATED AT 24.4CM DISTAL TO THE STRAIN RELIEF. BOTH SECTIONS OF THE HYPOTUBE BREAK WERE KINKED. NO ISSUES WERE NOTED WITH THE PROFILES OF THE CRIMPED STENT BALLOON AND TIP. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL AND/ OR PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT.(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE A CATHETER SHAFT FRACTURE OCCURRED. THE LESION WAS LOCATED AT THE RIGHT CORONARY ARTERY. A 2.75X16MM PROMUS ELEMENT DRUG ELUTING STENT WAS ADVANCED AND BREAKS ON THE SHAFT WERE NOTICED. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO REPORTED PATIENT COMPLICATIONS AND THE PATIENT STATUS IS STABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE A CATHETER SHAFT FRACTURE OCCURRED.THE LESION WAS LOCATED AT THE RIGHT CORONARY ARTERY. A 2.75X16MM PROMUS ELEMENT DRUG ELUTING STENT WAS ADVANCED AND BREAKS ON THE SHAFT WERE NOTICED. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO REPORTED PATIENT COMPLICATIONS AND THE PATIENT STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222811 PROMUS ELEMENT ? STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911316270 15752614

Patients

Seq Age Sex Outcome Treatment
1