ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-02434
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- April 19, 2013
- Report Date
- April 26, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED ¿ EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE ¿INSTRUMENT ERROR¿ ALERT WAS DISPLAYED; AND WHEN TESTED ON THE GEN04 GENERATOR, AN ERROR CODE 5 (INSTRUMENT ERROR) WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE TO STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 OR INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS ¿REPLACE INSTRUMENT¿ WITH THE GEN11 LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE. THE ¿RELAX PRESSURE ON BLADE; REACTIVE INSTRUMENT TO CONTINUE¿ YELLOW MESSAGE SCREEN IS ADVISING THAT THE INSTRUMENT HAS BEEN LOADED TO THE POINT WHERE OUTPUT HAS STOPPED. THE USER NEEDS TO RELAX PRESSURE ON THE INSTRUMENT OR REPOSITION SO THERE IS LESS TISSUE IN THE JAWS. RELEASE THE ACTIVATION SWITCH AND REACTIVATE THE INSTRUMENT TO CONTINUE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE THE GEN11 WAS GIVEN THE ERROR CODE RELEASE PRESSURE AS THE SURGEON WAS GOING TO THE OMENTUM. WHEN THE ERROR OCCURRED THE ACTIVE BLADE OF THE INSTRUMENT WAS BROKEN THE PIECE THAT WAS BROKEN OFF IS REMOVED FROM THE PATIENT. THEY OPENED A NEW DEVICE. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 221871 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |