FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3120612 · Received May 20, 2013

Report

Report Number
1416980-2013-12928
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
May 5, 2013
Report Date
May 6, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TESTING, AND CLAMP FUNCTION TESTING WERE PERFORMED ON THE DEVICE WITH NO ISSUES NOTED. THE DEVICE WAS RUN ON A LAB HOMECHOICE MACHINE WITH NO ISSUES NOTED. THE REPORTED PROBLEM WAS NOT CONFIRMED AND THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER?S TECHNICAL SERVICE CENTER THAT A HOME PATIENT (HP) HAD RECEIVED A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE) ON THE HOMECHOICE (HC) MACHINE DURING DWELL 2. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO START OVER WITH ALL NEW SUPPLIES AND ASSISTED WITH ENDING THE THERAPY. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222030 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 52 YR HOMECHOICE