FDA Adverse Event Injury Summary report: N

D/M 10 IN SS BROAD COMP PLATE

MDR report key: 3120494 · Received May 20, 2013

Report

Report Number
0002249697-2013-01717
Event Type
Injury
Date Received
May 20, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
HRS
PMA / PMN Number
K922525
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING CRACK/FRACTURE INVOLVING A DALL MILE PLATE WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION AND RESULTS INDICATE THAT FRACTURE OCCURRED THROUGH THE FOURTH FIXATION HOLE FROM ONE END OF THE PLATE. DIMENSIONAL INSPECTION NOT PERFORMED BECAUSE THERE IS NO INDICATION OF A DIMENSIONAL ISSUE. FUNCTIONAL INSPECTION NOT PERFORMED BECAUSE THE CLINICAL SITUATION CANNOT BE REPLICATED. THE MATERIAL ANALYSIS REPORT CONCLUDED THAT THE COMPRESSION PLATE FRACTURED THROUGH ONE OF THE CENTRAL FIXATION HOLES IN FATIGUE. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE DEVICE FEATURES EVALUATED. A MEDICAL REVIEW BY A CLINICAL CONSULTANT INDICATED THAT THE ROOT CAUSE FOR DEVELOPMENT OF THIS OVERLOAD CONDITION ARE PATIENT-RELATED FACTORS REGARDING DELAYED FRACTURE HEALING. ALL DEVICES IN THE REPORTED LOT WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THERE HAVE BEEN NO OTHER EVENTS FOR THE REPORTED LOT. THE ROOT CAUSE OF THE EVENT WAS DETERMINED TO BE THAT THE PLATE FRACTURED IN FATIGUE DUE TO AN OVERLOAD CONDITION RELATED TO DELAYED FRACTURE HEALING. THERE IS NO EVIDENCE THAT THE EVENT WAS MANUFACTURING RELATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PER PHYSICIAN AMBULATING. FRACTURE OF PLATE OCCURRED BETWEEN HOLES #5 AND 6 AT LEVEL OF THE ORIGINAL FRACTURE SITE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PER PHYSICIAN AMBULATING. FRACTURE OF PLATE OCCURRED BETWEEN HOLES #5 AND 6 AT LEVEL OF THE ORIGINAL FRACTURE SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221981 D/M 10 IN SS BROAD COMP PLATE IMPLANT HRS STRYKER ORTHOPAEDICS-MAHWAH 33004801

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| O| R