HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-12895
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- May 1, 2013
- Report Date
- May 1, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. USE ERROR AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE" WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE WARNS THE USER NOT TO REPLACE EMPTY SOLUTION BAGS OR RECONNECT DISCONNECTED SOLUTION BAGS DURING THERAPY. IT ALSO WARNS THE USER THAT POSSIBLE CONTAMINATION OF THE FLUID OR FLUID PATHWAYS CAN RESULT IF DISPOSABLES ARE REUSED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING TROUBLESHOOTING FOR ANOTHER ISSUE, IT WAS FOUND THAT THE HOME PATIENT (HP) HAD MADE AN USE ERROR WHILE USING A HOMECHOICE (HC) MACHINE WITH A CASSETTE. THE HP HAD STEPPED ON THE PATIENT LINE, CAUSING THE PATIENT LINE TO DISCONNECT FROM THE TRANSFER SET. THE HP THEN RECONNECTED AND WANTED TO BYPASS TO FILL. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO END THERAPY. THE TSR EXPLAINED THAT IF THE TRANSFER SET DISCONNECTS FROM THE PATIENT LINE SHE SHOULD NOT RECONNECT AND SHOULD END THERAPY AND START OVER WITH NEW SUPPLIES OR DO A MANUAL EXCHANGE INSTEAD. THE HP'S NURSE LATER STATED THAT THE EVENT WAS CAUSED BY USE ERROR AND NO ALLEGATIONS WERE MADE AGAINST ANY BAXTER PRODUCTS. THE HP WAS ADMINISTERED PROPHYLACTIC ANTIBIOTICS AND WAS CONTINUING THERAPY. THERE WERE NO ADVERSE EFFECTS REPORTED IN RELATION TO THIS EVENT. THIS IS THE SAME PATIENT AS (B)(4). ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222300 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | HOMECHOICE| TRANSFER SET| UNKNOWN DIALYSIS SOLUTION |