OT ULTRAMINI METER

MDR report key: 3119777 · Received May 19, 2013

Report

Report Number
3008382007-2013-12123
Event Type
Injury
Date Received
May 19, 2013
Report Date
April 30, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K061118
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING. THE ALLEGED ISSUE WAS NOT OBSERVED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRAMINI METER READ INACCURATELY ERRATIC. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ONE MORNING IN (B)(6) 2013. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿225, 103, 164 AND 218 MG/DL¿ WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF (B)(4). THE PATIENT MANAGES HER DIABETES WITH HUMALOG INSULIN AND HUMULIN INSULIN. THE PATIENT CONTINUED TO ADMINISTER HER USUAL DOSE OF MEDICATIONS. THE PATIENT CLAIMED SHE FELT SYMPTOMS OF SWEATY, TIRED AND DIZZY. THE PATIENT REPORTEDLY OBTAINED A RESULT OF ¿65 MG/DL¿ WITH ANOTHER METER. THE PATIENT TOOK FOOD AND/ OR DRANK A BEVERAGE AS TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO PERFORM QUALITY CONTROL TESTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY MAY HAVE DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220648 OT ULTRAMINI METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3331203

Patients

Seq Age Sex Outcome Treatment
1 70 YR Life Threatening| R