FDA Adverse Event Malfunction Summary report: N

DISTAL TARGETING DEVICE GAMMA3 DISTAL TARGETING R1.5

MDR report key: 3119032 · Received May 19, 2013

Report

Report Number
0009610622-2013-00267
Event Type
Malfunction
Date Received
May 19, 2013
Date of Event
April 23, 2013
Report Date
April 23, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-KIEL
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED ISSUE WAS NOT CONFIRMED. THE EVENT DESCRIPTION DEFINED THE DISTAL TARGETING DEVICE (DTD) TO BE THE PRIMARY PRODUCT. NO ASSOCIATED PRODUCTS WERE REPORTED. AN INVESTIGATION AND A REVIEW OF THE DHR WERE NOT POSSIBLE BECAUSE THE PRODUCT AND LOT CODE WERE NOT PROVIDED; THE ROOT CAUSE COULD NOT BE DETERMINED. A POSSIBLE ROOT CAUSE FOR THE FUNCTIONAL PROBLEM COULD BE THAT THE WASHERS INSIDE THE FIXATION MECHANISM WERE ROTATED AND CAUSED THE NON-FUNCTION OF THE FIXATION. A CHANGE REQUEST WAS PERFORMED TO REPLACE THE WASHERS TO AN ELASTIC ELEMENT WITH THE SAME GEOMETRIES. THE CHANGE REQUEST BECAME EFFECTIVE IN MAY 2013. IF THE COMPLAINED DTD WAS MANUFACTURED PRIOR TO THE CHANGE REQUEST COULD NOT BE DETERMINED BECAUSE THE MANUFACTURING DATE IS UNKNOWN. A MORE PRECISE EVALUATION WAS NOT POSSIBLE DUE TO MISSING INFORMATION. NO DISCREPANCIES WERE DETECTED DURING RISK ANALYSIS REVIEW. DEVICE WILL NOT BE RETURNED.

Description of Event or Problem · 1

DURING T2 RECON NAIL W/S, WHEN CONNECTING THE DISTAL TARGETING ARM AND T2 RECON TARGET DEVICE, FIXATION BOLT COULD NOT INSERT TO THE DEVICE HOLE.

Description of Event or Problem · 1

DURING T2 RECON NAIL W/S, WHEN CONNECTING THE DISTAL TARGETING ARM AND T2 RECON TARGET DEVICE, FIXATION BOLT COULD NOT INSERT TO THE DEVICE HOLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221235 DISTAL TARGETING DEVICE GAMMA3 DISTAL TARGETING R1.5 INSTRUMENT LXH STRYKER OSTEOSYNTHESIS-KIEL UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other