FDA Adverse Event Summary report: N

60" EVOLUTION MEDIUM STERILIZER

MDR report key: 3118949 · Received May 19, 2013

Report

Report Number
3005899764-2013-00043
Date Received
May 19, 2013
Date of Event
April 18, 2013
Report Date
May 17, 2013
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED AT THE FACILITY, INSPECTED THE UNIT AND FOUND THE STEAM PRESSURE REGULATOR VALVE (PRV) BROKEN. THE STEAM PRV IS COMPOSED OF TWO PARTS: THE STEM ASSEMBLY AND SYLPHON. THE STEM ASSEMBLY CONTROLS THE AMOUNT OF STEAM TO THE UNIT, AND IS CLOSED ONCE THE CORRECT AMOUNT OF PRESSURE IS MET. THE TECHNICIAN FOUND THE STEM ASSEMBLY AND SYLPHON BOTH BROKEN INTO TWO PARTS. THE TECHNICIAN INSTALLED A NEW STEAM PRV INTO THE UNIT AND CONFIRMED THE UNIT TO BE OPERATIONAL. THE UNIT WAS INSTALLED IN (B)(6) 2010 AND IS CURRENTLY UNDER A STERIS SERVICE CONTRACT. THE LAST PREVENTATIVE MAINTENANCE FOR THE UNIT WAS COMPLETED ON (B)(6) 2013. DURING THIS TIME, THE TECHNICIAN REPLACED THE STEAM PRV, AS INSTRUCTED IN THE PREVENTATIVE MAINTENANCE CHECK LIST, AND CONFIRMED THE UNIT TO BE OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THEIR UNIT WAS LEAKING STEAM. NO INJURIES OR PROCEDURAL DELAYS/CANCELLATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220617 60" EVOLUTION MEDIUM STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1