FDA Adverse Event
Other
Summary report: N
NON21430CNS
MDR report key: 3115281
·
Received April 29, 2013
Report
- Report Number
- 3005841380-2013-00003
- Event Type
- Other
- Date Received
- April 29, 2013
- Date of Event
- May 24, 2012
- Report Date
- April 25, 2013
- Manufacturer
- CLEARCOUNT MEDICAL SOLUTIONS
- Product Code
- GDY
- PMA / PMN Number
- K093557
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS MDR IS BEING FILED AT THE BEHEST OF FDA IN RESPONSE TO MW REPORT (B)(4). THIS COMPLAINT WAS REC'D BY CLEARCOUNT MEDICAL SOLUTIONS ON (B)(4) 2012, INVESTIGATED AND FOUND TO NOT REQUIRE AN MDR. THE ORIGINAL COMPLAINT STATED THAT A TAG WOULD NOT READ,WHICH WAS FOUND TO BE TRUE AFTER INVESTIGATION. THE ROOT CAUSE OF THE FAILURE OF THE TAG WAS FOUND TO BE A FAILED SOLDER JOINT. A REPORT OF THE INVESTIGATION WAS ISSUED TO (B)(6) BY (B)(6), VP QARA.
Description of Event or Problem · 1
ITEM 3 FROM REPORT (B)(4), RFID NOT DETECTED WHEN WANDED. NO PTS WERE HARMED. THIS EVENT WAS REPORTED TO CLEARCOUNT MEDICAL SOLUTIONS ON (B)(6) 2012 AS A FAILED TAG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 185349 | NON21430CNS | 4X4 GAUZE SPONGE WITH FRID TAG | GDY | CLEARCOUNT MEDICAL SOLUTIONS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |