ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06447
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Report Date
- April 17, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED DATES FROM (B)(6) 2013 WITH NO EVIDENCE OF A REBOOT. DURING INVESTIGATION OF THE PUMP, NO DAMAGE OR INTERNAL MOISTURE WAS FOUND TO THE BATTERY COMPARTMENT. THE BATTERY CAP WAS ABLE TO TIGHTEN ON TO THE PUMP. THE PUMP¿S DISPLAY SCREEN WAS BLANK WHEN ATTEMPTING TO POWER ON THE PUMP. THE POWER ISSUE COULD NOT BE FURTHER TESTED DUE TO THE BLANK DISPLAY. THE DISPLAY SCREEN WAS FOUND TO BE CRACKED. A TEST SCREEN WAS INSTALLED AND DISPLAYED PROPERLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. IT WAS REPORTED THAT THE PUMP HAD NO POWER AND THE DISPLAY SCREEN APPEARED TO BE SHATTERED FROM THE INSIDE WHEN THE PUMP WAS REMOVED FROM A BAG HOLDING SUPPLIES DURING A SPORTS ACTIVITY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 215575 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |