ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-02340
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Date of Event
- March 28, 2013
- Report Date
- April 1, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE A WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED. THIS BLADE TIP PORTION MAY HAVE BROKEN OFF OF THE DEVICE DURING TRANSPORT TO OUR ANALYSIS SITE. THE REPORTED COMPLAINT WAS FOR AN ERROR 5 WAS DISPLAYED DURING USE. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. WHEN A BLADE IS DAMAGED, IT MAY NOT COMPLETE THE PRE-RUN TESTING, WILL DISPLAY AN ERROR CODE AND WILL KEEP REVERTING BACK TO STANDBY MODE. A PROBABLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. IF THE HARMONIC SYSTEM SENSES AN INSTRUMENT FAULT DURING USE, AN AUDIBLE ALARM (TONE WITH LONG PULSES) WILL SOUND AND A VISUAL ALARM INDICATOR WILL APPEAR ON THE CONTROL PANEL. THE GENERATOR SYSTEM WILL REVERT TO STANDBY MODE, WHEN THE AUDIBLE ALARM (SOLID TONE) AND VISUAL ALARM INDICATOR APPEAR. DEVICE B WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED WITH A GENERATOR AND WAS FUNCTIONAL, IN ADDITION, THE DEVICE ACTIVATED WITH BOTH MAX AND MIN BUTTONS. THE TISSUE PAD OF THE DEVICE WAS IN GOOD PHYSICAL CONDITION. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. IT COULD NOT BE DETERMINED WHY REPORTEDLY AN ERROR 5 WAS DISPLAYED DURING USE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. DEVICE B: BATCH J92Z8J, MFG DATE 12/1/2012, EXP DATE 11/12/2017.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, AN ERROR 5 WAS DISPLAYED DURING USE. THE 2ND DEVICE WAS USED, BUT THE SAME EVENT OCCURRED. THE 3RD DEVICE AND ANOTHER HAND PIECE WERE USED, THE 3RD DEVICE WAS ACTIVATED. THE 3RD DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE BLADE OF THE 1ST DEVICE WAS BROKEN OFF WHEN A CLINICAL ENGINEER CHECKED IT AFTER OPERATION. NO PIECES FELL INTO THE PATIENT. TWO DEVICES WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214304 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | J92Z8D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |