FDA Adverse Event Malfunction Summary report: N

AXON COMPUTER NOTEBOOK

MDR report key: 3113314 · Received May 15, 2013

Report

Report Number
1030489-2013-01700
Event Type
Malfunction
Date Received
May 15, 2013
Report Date
April 18, 2013
Manufacturer
MEDTRONIC XOMED, INC
Product Code
GWF
PMA / PMN Number
K061639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERIAL NUMBER: (B)(4), LOT NUMBER: 68068853H ,THE SERIAL NUMBER NOTED WAS PROVIDED BY THE CUSTOMER, BUT IS NOT VALID. DEVICE RECEIVED ON JULY 17 2013, FOR EVALUATION. INFORMATION RECEIVED ON JULY 22, 2013. EVALUATION SUMMARY: PRODUCT ANALYSIS FOR THE 945NCCPU NOTEBOOK (S/N (B)(4)) FOUND THAT THE HARD DRIVE WAS CORRUPT. THE HARD DRIVE WAS RE-IMAGED AND THE LAPTOP TESTED. THE LAPTOP PASSED MANUFACTURING SPECIFICATIONS. NOT A SINGLE USE DEVICE.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. (B)(4). THE DEVICE WAS NOT RETURNED AND THEREFORE NO EVALUATION COULD BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE A COMPUTER NOTEBOOK WAS RUNNING SLOW AND LOCKED UP ON SURGEON DIRECTED. THERE WAS NO PATIENT IMPACT.