FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 3113243
·
Received May 8, 2013
Report
- Report Number
- 1720753-2013-05843
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- May 2, 2013
- Report Date
- May 8, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE HEATER BOARDS WERE REPLACED DURING THE SVC CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYS DISPLAYED HEATER ERRORS AND WOULD NOT PERFORM FLUOROSCOPY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202339 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |