FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3113235 · Received May 8, 2013

Report

Report Number
1720753-2013-05831
Event Type
Malfunction
Date Received
May 8, 2013
Date of Event
May 1, 2013
Report Date
May 8, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE BACKPLANE AND HIGH VOLTAGE CABLE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A KV ERROR MESSAGE AND WOULD NOT PRODUCE AN IMAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202465 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1