FDA Adverse Event
Malfunction
Summary report: N
LIFECARE PCA 3 V5. 06
MDR report key: 3113207
·
Received May 8, 2013
Report
- Report Number
- 9615050-2013-01092
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 10, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- MEA
- PMA / PMN Number
- K042800
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THIS WAS DUE TO WORN THREADS ON THE HALF NUT. AS INDICATED, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200287 | LIFECARE PCA 3 V5. 06 | 80MEA | MEA | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |