FDA Adverse Event
Malfunction
Summary report: N
DUAL COOLER/HEATER
MDR report key: 3113168
·
Received May 9, 2013
Report
- Report Number
- 1828100-2013-00444
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- April 17, 2013
- Report Date
- April 18, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- PREAMEND
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) CHECKED THE R1-R3 RESISTOR BANKS: R1 = 35 OHMS, R2 = 225 MEG OHMS, R3 = 50 OHMS. THE FSR ALSO NOTED THE UNIT WOULD PUMP IN ALL HEATING MODES AND THE COOL 3 AND COOL 4 SETTINGS, BUT THE UNIT WOULD NOT PUMP IN COOL #1 NOR COOL #2 SETTINGS. THE FSR ADVISED THE CUSTOMER THAT SPEED CONTROL RESISTORS R2 AND R3 NEED REPLACEMENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT WOULD NOT PUMP WATER IN THE "RECIRCULATION" MODE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202716 | DUAL COOLER/HEATER | COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 11160 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |