FDA Adverse Event Malfunction Summary report: N

LC PCA REFURB 802.11

MDR report key: 3113157 · Received May 8, 2013

Report

Report Number
9615050-2013-01111
Event Type
Malfunction
Date Received
May 8, 2013
Date of Event
April 2, 2013
Report Date
April 17, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
MEA
PMA / PMN Number
K042800
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THE DEVICE WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPARTMENT WITH A NOTE THAT STATED, "FAILURE MESSAGE KEEPS COMING UP AFTER ABOUT 30 MINUTES STARTING INFUSION." NO SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE PROVIDED. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202424 LC PCA REFURB 802.11 80MEA MEA HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK