FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3113065 · Received May 9, 2013

Report

Report Number
1720753-2013-05881
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 30, 2013
Report Date
May 9, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SCL)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE SINGLE BOARD COMPUTER AND REFORMATTED THE HARD DRIVE AND RELOADED SOFTWARE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM WOULD NOT COMPLETED BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204716 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SCL) 9800

Patients

Seq Age Sex Outcome Treatment
1