FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3112950 · Received May 14, 2013

Report

Report Number
3008382007-2013-11542
Event Type
Malfunction
Date Received
May 14, 2013
Report Date
April 23, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (6/17/2013) THE PATIENT¿S TEST STRIPS HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT PASSED TESTING. THE TEST STRIPS WERE EVALUATED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED; HOWEVER, A SECONDARY ISSUE WAS NOTED WHERE ERROR 4 WAS OBSERVED DURING CONTROL SOLUTION TESTS. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIO IQ METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "167 AND 187 MG/DL" WITH THE SUBJECT METER AND "109 AND 149 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER, RESPECTIVELY. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO METER ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212561 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3410688

Patients

Seq Age Sex Outcome Treatment
1 65 YR