FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3112904 · Received May 14, 2013

Report

Report Number
3008382007-2013-11486
Event Type
Malfunction
Date Received
May 14, 2013
Report Date
April 19, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (07/26/2013) THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE HIGH RESULTS COMPARED TO ANOTHER DEVICE. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF 183 MG/DL WITH THE SUBJECT METER AND 108 MG/DL ON ANOTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212138 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3318425

Patients

Seq Age Sex Outcome Treatment
1