VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2013-00383
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN THE FEEDSET TUBE WHERE IT IS INSERTED INTO THE CHAMBER. THE SURFACE OF THE BREAK IS ROUGH AND NOT SMOOTHLY CUT. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 120719. CONCLUSION: THE DAMAGE APPEARED TO BE DUE TO THE FEEDSET TUBE BEING PULLED, AWAY FROM THE CHAMBER, POSSIBLY FROM THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY ARE RELEASED FOR DISTRIBUTION AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT A WATER LEAK OCCURRED AT THE FEEDSET OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213462 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 120719 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |