FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3112830 · Received May 14, 2013

Report

Report Number
2955842-2013-01666
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
March 25, 2013
Report Date
April 17, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE INSTRUMENT WAS NOT CUTTING. THE INSTRUMENT GRIP CABLE WAS FOUND BROKEN AND DERAILED AT THE DISTAL IDLER PULLEY, RESULTING IN THE INSTRUMENT NOT BEING ABLE TO CUT. AN ADDITIONAL OBSERVATION FOUND DAMAGE ON THE SURFACE AND THE EDGE OF THE IDLER PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI SACROLPOPEXY PROCEDURE THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT WAS NOT CUTTING. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212778 LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-02 M10111209 937

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES