FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT

MDR report key: 3112816 · Received May 14, 2013

Report

Report Number
9611451-2013-00386
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 15, 2013
Report Date
April 16, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) (B)(4) AND VISUALLY INSPECTED. RESULTS: A HOLE WAS FOUND IN THE EVAQUA (EXPIRATORY) LIMB OF THE COMPLAINT BREATHING CIRCUIT, APPROXIMATELY 49CM AWAY FROM THE PATIENT END CONNECTOR. CONCLUSION: BASED ON INSPECTION OF THE HOLE DAMAGE, IT IS LIKELY THAT THE EVAQUA EXPIRATORY LIMB WAS PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO BEING RELEASED FOR DISTRIBUTION. (B)(4). ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS DAMAGED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. (B)(4). THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT RT340 DEVICE IS CURRENTLY IN TRANSIT TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE SPECIALIST THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. A PIN HOLE WAS FOUND ON THE EXPIRATORY LIMB. THIS WAS FOUND PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE SPECIALIST THAT AN RT340 ADULT BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. A PIN HOLE WAS FOUND ON THE EXPIRATORY LIMB. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213613 ADULT DUAL-HEATED WITH EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 120918

Patients

Seq Age Sex Outcome Treatment
1