MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2013-01660
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 3, 2013
- Report Date
- April 16, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE COMPLAINT THE ORANGE SLEEVE WAS CRACKED WAS CONFIRMED. FOR CLARIFICATION, THE TUBE EXTENSION WAS CRACKED AT THE PROXIMAL CLEVIS INTERFACE. THE CLEVIS WAS DISLODGED FROM THE TUBE EXTENSION AS A RESULT. ENGINEERING CONCLUDED THE TUBE EXTENSION LIKELY BROKE DUE TO EXCESSIVE SIDE LOADING OR OTHER MISHANDLING. ADDITIONAL DAMAGE FOUND WAS DEEP SCRATCHES EXHIBITING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH ON THE DISTAL END OF THE MAIN TUBE. ENGINEERING CONCLUDED THE DAMAGE MAY BE DUE TO MISHANDLING. ELECTRICAL CONTINUITY PASSED. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DEEP SCRATCHES EXHIBITING LIGHT MATERIAL REMOVAL, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI SI MYOMECTOMY PROCEDURE THE ORANGE SLEEVE WAS CRACKED ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT AND WAS UNABLE TO BE REMOVED FROM THE TROCAR. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213029 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-10 | M10121206 499 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES ESU |