FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3112660 · Received May 14, 2013

Report

Report Number
3004209178-2013-93458
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 18, 2013
Report Date
April 18, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM EXCESSIVE NO DELIVERY ALARMS TEST DUE TO LOOSE DRIVE SUPPORT DISK. NO BURN WIRE OR OTHER COMPONENTS NOTED. UNIT HAD BROKEN RESERVOIR TUBE LIP AND BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT MULTIPLE NO DELIVERY ALARMS AND SHE STATED THAT LOOKS LIKE A WIRE OR SOMETHING INSIDE OF THE INSULIN PUMP CAUGHT ON FIRE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212167 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 44 YR