FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112650 · Received May 14, 2013

Report

Report Number
3004209178-2013-93421
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
February 17, 2013
Report Date
April 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARM MOTOR ERROR DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT WAS RECEIVED WITH CRACKED DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP IS CRACKED AND THE INSULIN IS SQUIRTING OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213416 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 63 YR