FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 3112648 · Received May 14, 2013

Report

Report Number
3004209178-2013-93419
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 17, 2013
Report Date
April 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE HAD HIGH BLOOD GLUCOSE DUE TO THE RESERVOIR WAS LEAKING INSULIN INTO RESERVOIR COMPARTMENT DURING NORMAL USE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212191 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A

Patients

Seq Age Sex Outcome Treatment
1 27 YR