FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3112646 · Received May 14, 2013

Report

Report Number
3004209178-2013-93450
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 15, 2013
Report Date
April 16, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK AND SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS LOOSE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212753 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 23 YR