FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112643 · Received May 14, 2013

Report

Report Number
3004209178-2013-93444
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 9, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH DAMAGE BATTERY CAP COIN SLOT, CRACKED RESERVOIR TUBE LIP, UNIT FAILED DISPLACEMENT TEST WITH UNITS LEFT ON STATUS SCREEN, AND MOTOR ERROR ALARM DURING REWIND DUE TO OUT OF PHASE MOTOR ENCODER SIGNAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING NORMAL USED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212701 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 66 YR