FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112637 · Received May 14, 2013

Report

Report Number
3004209178-2013-93437
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
March 7, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARM DURING THE PRIME TEST DUE TO LOOSE MOTOR DRIVE SUPPORT DISK. UNIT HAD MISSING END CAP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD INSULIN SQUIRTING OUT AND ALARMING DURING MANUAL PRIME. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212670 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWL

Patients

Seq Age Sex Outcome Treatment
1