FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112636 · Received May 14, 2013

Report

Report Number
3004209178-2013-93436
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
March 7, 2013
Report Date
May 9, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARM MOTOR ERROR DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS SQUIRTING THE INSULIN OUT AND ALARMING MOTOR ERROR DURING NORMAL USED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212294 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWS

Patients

Seq Age Sex Outcome Treatment
1