FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112634 · Received May 14, 2013

Report

Report Number
3004209178-2013-93433
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
February 6, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARM MOTOR ERROR DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT WAS RECEIVED WITH MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT THAT SHE HAD HIGH BLOOD GLUCOSE AND THE INSULIN PUMP IS ALARMING MOTOR ERROR AND SQUIRTING THE INSULIN OUT DURING THE MANUAL PRIME. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212669 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 55 YR