RESERVOIR 1.8ML
Report
- Report Number
- 3004209178-2013-93478
- Event Type
- Injury
- Date Received
- May 14, 2013
- Date of Event
- April 17, 2013
- Report Date
- April 19, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.
CUSTOMER CALLED TO REPORT HOSPITALIZATION DUE TO DIABETIC KETOACIDOSIS. INSULIN HAD LEAKED INTO THE RESERVOIR COMPARTMENT. CUSTOMER'S BLOOD GLUCOSE READING IS 300 MG/DL. CUSTOMER HAS TREATED WITH INSULIN. INSULIN HAS LEAKED PAST THE SECOND O-RING. CUSTOMER STATED THAT THE RESERVOIR LEAKED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212202 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Hospitalization |