RESERVOIR 1.8ML

MDR report key: 3112604 · Received May 14, 2013

Report

Report Number
3004209178-2013-93474
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 19, 2013
Report Date
April 19, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

INSPECTED THREE RANDOMLY SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY. INSPECTED TWO OPENED AND USED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED BY CUSTOMER'S MOTHER THAT THE RESERVOIR LEAKED. CALLER STATED THAT THE CUSTOMER BELIEVES THE BLOOD GLUCOSE READING IS HIGH. THE BLOOD GLUCOSE READING IS 62 MG/DL. THE LEAK OCCURRED DURING THE NIGHT AND THE NEXT DAY. THE RESERVOIR LEAKED INTO THE RESERVOIR COMPARTMENT. THE RESERVOIR LEAKED PAST THE SECOND O-RING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213325 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 12 YR