FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3112603 · Received May 14, 2013

Report

Report Number
3004209178-2013-93473
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 19, 2013
Report Date
April 19, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. MOTOR PASSED MOTOR TEST.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMS MOTOR ERROR. CUSTOMER'S BLOOD GLUCOSE READING IS 164 MG/DL. ASSISTED CUSTOMER WITH CLEARING THE ALARM. CUSTOMER IS ABLE TO REWIND THE INSULIN PUMP. DISPLACEMENT TEST PASSED. CUSTOMER STATED THAT MOTOR ERROR ALARM OCCURS DURING BOLUS DELIVERY. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212607 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 49 YR