FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3112602
·
Received May 14, 2013
Report
- Report Number
- 3004209178-2013-93477
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 1, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED PRIME ALARM DURING THE PRIME TEST DUE TO LOOSE/PROTRUDED SUPPORT DISK. MISSING END CAP STICKER NOTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED PRIME ALARM AND THE INSULIN SQUIRTED OUT OF THE INFUSION SET. CUSTOMER'S BLOOD GLUCOSE READING IS 220 MG/DL. THE INSULIN IS SQUIRTING OUT DURING THE MANUAL PRIME PROCESS. THE DRIVE SUPPORT DISK IS PROTRUDED. ADVISED CUSTOMER TO DISCONTINUE THE USE OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 212201 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |